Description
Coating Thickness Meter TIME2600
Order Number:
TIME2600: with F1, N1 probe
TIME2600F: with F1 probe
TIME2600N: with N1 probe
TIME2600F4: with F400 probe
TIME2600F19: with F1/90 probe
TIME2600CN: with CN02 probe
Product Introduction:
TIME2600 Coating Thickness Gauge can nondestructively measure the thickness of non-magnetic coatings (such as zinc, aluminum chrome, copper, rubber, paint and so on) on ferrous metals (such as steel, iron, alloyed metals, hard magnetic steel and so on). It can also measure the thickness of non-conducting coatings (such as rubber, paint, plastic, anode oxide film and so on) on the non-ferrous materials (such as copper, aluminum, zinc, tin and so on). It can fast, exactly and nondestructively measure the thickness of coating and plating.
Features
1. Two measuring methods: magnetic induction (F) and eddy current (N). When F probe is equipped, it can measure the non-magnetic coating in magnetism substrate, when N probe is equipped, it measure the non-conductive coating on non-magnetism substrate.
2. 6 types of probes are available for various applications:F400, F1, F1/90, F10, N1, CN02
3. 2 measurement modes: continuous / single
4. 5 statistical ways: Mean values / Max. values / Min. values / testing numbers. (No.) / standard deviations (S.DeV)
5. 640 measured values can be stored
6. Integrated with printer for convenient use
7. Low battery indication
8. 2 switch off modes: manual and auto
Technical Specification
1. Standards: DIN, ISO, ASTM, BS
2. Calibration: Zero and foil calibration
3. Data memory: 640 readings
4. Statistics: Number of measurements, mean, standard deviation, maximum and minimum of 3000 readings
5. Data memory: 640 readings
6. Limits: Adjustable with alarm
7. Interface: RS-232
8. Working temperature: 0~40℃
9. Power supply: NiMH Rechargeable batteries 1.25V
10. Dimensions: 230 x 86 x 47mm
11. Weight: Approx. 530g
Probe model | F400 | F1 | F1/90º | F10 | N1 | CN02 | |||
Operating principle | Magnetic induction | Eddy current | |||||||
Measuring range (µm) | 0-400 | 0-1250 | 0-10000 | 0 to 1250μm 0 to 40μm (for chrome plate on copper) |
10~200 | ||||
Low range resolution (µm) | 0.1 | 0.1 | 10 | 0.1 | 1 | ||||
Accuracy | One-point calibration (µm) | ±(3%H+1) | ±(3%H+10) | ±(3%H+1.5) | ±(3%H+1) | ||||
Two-point calibration (µm) | ±[(1~3)H%+0.7] | ±[(1~3)H%+1] | ±[(1~3)%H+10] | ±[(1~3)%H+1.5] | – | ||||
Measuring conditions | Min curvature of the min area (mm) | Convex | 1 | 1.5 | Flatten | 10 | 3 | Flatten | |
Diameter of the min area (mm) | φ3 | φ7 | φ7 | φ40 | φ5 | φ7 | |||
Critical thickness of substrate (mm) | 0.2 | 0.5 | 0.5 | 2 | 0.3 | unlimited |